Overview
Our Support team has received a number of calls from offices that are unable to acquire radiographs using Carestream sensors. Our developers have investigated and found the cause to be a recently released Windows Update installed on September 8, 2020. The Update numbers are KB4571756 and KB4576478. There is a third update from October 12, 2020 which is KB4579311. There is a fourth update from November 11, 2020 which is KB458678. The fifth update is KB4592438. Sixth was installed on 1/14 that update is KB4598242. Seventh update installed on 2/15 is KB4601319.
If you are using Carestream sensors, we would recommend pausing Windows Updates until these issues have been resolved. Reboot your computer and test.
If the updates have already been installed on your computers, contact your IT Professional to have them remove updates KB4571756, KB4576478, KB4579311, KB4586781, KB4592438, KB4598242, KB4601319 and from all operatory computers. Reboot Computer and test the sensor
Detailed Instructions
If you are running version 12.355x of TDO or above please follow these steps. This is a computer specific fix and will need to be run on all operatory computers using a Carestream sensor.




Article ID: 848
Created: December 29, 2020
Last Updated: November 20, 2025
Author: TDO KB [web_admin@tdo4endo.com]
Online URL: https://kb.tdo4endo.com/article.php?id=848