Practices using Schick sensors for digital radiography have two supported integration options with TDO: the Sidexis 4.4 Bridge and the SiTwain driver. Both are separate installs that must be set up by Dentsply Sirona support before any TDO configuration can take place. For upgrade assistance, existing Schick users can visit dsgo.to/s4upgrade, or contact Dentsply Sirona Technical Support at 1 (800) 659-5977.
Note: The CDR Intraoral Twain driver previously used for Schick sensor acquisition is no longer supported. It is incompatible with Windows 11, and TDO Tech Support cannot assist with issues arising from that integration. Practices still on CDR Intraoral Twain should transition to one of the two supported options outlined in this article.
The Sidexis 4.4 Bridge (supported on TDO v12.463 and above) allows users to launch the Sidexis software from TDO, and images can then be copied and pasted into TDO Acquire to be saved directly to the patient's chart. The SiTwain driver (supported on TDO v12.485a and above) provides a direct acquisition path as an alternative to the full Sidexis application, though note that a valid Sidexis license is still required. Setup steps for both are covered in the sections below.
Requirements:
TDO Configuration:
In TDO, go to Setup > Site Configuration > Edit Control Table > Workstation Settings. Set the Digital Radiography settings like so:
That is all that is needed for setup! From TDO, you can then launch the SiTwain from Image Acquire> Tooth button
You will be able to take 1 X-ray at a time, but when pressing finish, the Twain UI will reopen to continue acquiring.
Requirements:
Installing Slida Configuration Tool
Note: As of 4/22/26, The SLIDA tool has been removed from the Sirona website. Due to this, we recommend the SiTwain driver option.




Steps to Bridge Sidexis with TDO 12:




In some computers, the patch could be \\[Server Name]\PDATA\siomin.sdx. If this is the case the INI should be like this: SidexisPathToSiomin=\\[Server Name]\PDATA\
SlidaPath=C:\Users\[Username]\Desktop\SLIDA_SDK\SidexisSlidaConfiguration\Bin

11.Close the Slida configuration
12. Launch TDO
13. Go to Control Table> Workstation and change the rad Software to “Sirona”, Radiography Path/Source C:\Program Files\Sirona\Sidexis4 or C:\Sidexis\
IMPORTANT: Please note, this is the path where Sidexis is installed in the computer, or where the SIFILEDB.ini is located. TDO uses this path to look for the executable file called Sidexis.exe or Sidexis4.exe.




Note: If Sidexis is left open and and another patient is selected in TDO, Sidexis will switch to the new patient
Troubleshooting

If this error pop ups when testing, please make sure the “Radiography Path” in “Workstation Settings” in TDO is pointing to where the SIFILEDB.ini is located

Article ID: 996
Created: April 18, 2025
Last Updated: April 23, 2026
Author: TDOKB2 [IT@tdo4endo.com]
Online URL: https://kb.tdo4endo.com/article.php?id=996